Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 73: (1 year) CAL VER /5700 DATE: 16-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 2.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 18 NUMBER OF LINES: 91 CONFIGURATION: Fluke 5700A ============================================================================= # # Source: # Fluke 70 Series service manual (PN 731034). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 73_DSP.bmp # 57_73.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F W 1.002 ASK+ K 1.003 MMFC C2 1.004 HEAD EQUIPMENT SETUP 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {DISPLAY TEST} 1.007 PICE 73_DSP 2.001 DISP Ambient temperature: 18C - 28C. 2.001 DISP Relative humidity: less than 80%. 2.001 DISP Allow the UUT to stabilize. 2.002 HEAD {RESISTANCE TESTS} 2.003 PIC 57_73 2.004 DISP Select the Ohms function. 2.005 HEAD RESISTANCE TESTS: {320 Ohm Range} 2.006 5700 320 100.0Z 1.2U CW 3.001 HEAD RESISTANCE TESTS: {3200 Ohm Range} 3.002 5700 3200 1000Z 11U CW 4.001 HEAD RESISTANCE TESTS: {32 kOhm Range} 4.002 5700 32 10.00kZ 0.11U CW 5.001 HEAD RESISTANCE TESTS: {320 kOhm Range} 5.002 5700 320 100.0kZ 1.1U 2W 6.001 HEAD RESISTANCE TESTS: {3.2 MOhm Range} 6.002 5700 3.2 1.000MZ 0.011U 2W 7.001 HEAD RESISTANCE TESTS: {32 MOhm Range} 7.002 5700 32 10.00MZ 0.31U 2W 8.001 HEAD {DC MILLIVOLT TEST} 8.002 DISP Select the 300mV DC Function. 8.003 HEAD DC MILLIVOLT TEST: {320mV Range} 8.004 5700 320 300.0mV 2.2U 2W 9.001 HEAD {DC VOLTAGE TEST} 9.002 DISP Select the V DC Function. 9.003 HEAD DC VOLTAGE TEST: {3.2V Range} 9.004 5700 3.2 2.700V 0.02U 2W 10.001 HEAD DC VOLTAGE TEST: {32V Range} 10.002 5700 32 27.00V 0.2U 2W 11.001 HEAD DC VOLTAGE TEST: {320V Range} 11.002 5700 320 270.0V 2.0U 2W 12.001 HEAD DC VOLTAGE TEST: {1000V Range} 12.002 5700 1000 1000V 9U 2W 13.001 HEAD {AC VOLTAGE TESTS} 13.002 DISP Select the V AC Function. 13.003 HEAD AC VOLTAGE TEST: {3.2V Range} 13.004 5700 3.2 2.700V 0.083U 100H 2W 14.001 5700 3.2 2.700V 0.083U 500H 2W 15.001 HEAD AC VOLTAGE TEST: {750V Range} 15.002 5700 750 750V 25U 100H 2W 16.001 5700 750 750V 25U 1000H 2W 17.001 HEAD {DC CURRENT TESTS} 17.002 DISP Select the A DC Function. 17.003 HEAD DC CURRENT TESTS: {10A Range} # 10A Range done at 1.90A instead of 2.00A (1.99999A) to allow for slew. 17.004 5700 10 1.90A 2% 0.02U 2W 18.001 END